G1362A Brechungsindexdetektor der Series 1100
The detector is designed for highest optical performance, GLP compliance and easy maintenance. It includes the following features:
- advanced temperature controlled detector optics ready to use within two hours of installation
- automatic zero and automatic purge combined with a recycle valve for automatic solvent recycling allow uninterrupted operation
- durable tungsten lamp with a life expectancy of 40,000 hours
- automatic light intensity control circuit to ensure the optimum performance of the optics
- integrated diagnostics for efficient troubleshooting
- built-in refractive index calibration
- front access to valves and capillaries for easy maintenance
Art.Nr.: G1362A_01
Produktbeschreibung
Spezifikationen
Type
|
Specification | Comments |
Detection type | Refractive Index | |
Refractive index range | 1.00 - 1.75 RIU, calibrated | |
Measurement range |
+/- 600 x 10^5 RIU
|
|
Optical zeroing | via set screw | |
Optics temperature
control
|
5 ° C above ambient to 55 ° C | |
Sample cell | volume 8uL maximum pressure 5 bar (0.5Mpa) maximum flow rate 5mL/minute | |
Valves |
Automatic purge and
automatic solvent recycle
|
|
Volumes | Inlet port to sample cell 62uL, inlet port to outlet port 590uL | |
Liquid contact materials |
316
stainless steel, teflon and quartz glass
|
|
pH range | 2.3 - 9.5 | |
Performance
specifications
|
Short term noise < +/- 2.5 x 10^9 RIU. Drift < 200 x 10^-9 RIU/hour | see note below this table |
Time programmable parameters | polarity, peak width | |
Detector zero |
automatic
zero before analysis
|
|
Control and data
evaluation
|
Parameter entry, signal display,on-line help and diagnostics with the Agilent 1200 Series Control Module. Optional PCMCIA card for method, sequence and logbook storage and transfer. Agilent ChemStation for LC PC based software for control and data evaluation. | |
Analog outputs | Recorder/integrator: 100 mV or 1 V, output range selectable, one output | |
Communications | Controller-area network (CAN), GPIB, RS-232C, LAN, APG Remote: ready, start, stop and shut-down signals | |
Safety and maintenance | Extensive diagnostics, error detection and display (through control module and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas | |
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-selectable limits and feedback messages. Electronic records of maintenance and errors. Automated operational qualification/performance verification (OQ/PV) | |
Housing | All materials recyclable. | |
Environment | 0 to 55 ° C constant temperature at <95% humidity (non-condensing) | |
Dimensions | 180 mm x 345 mm x 435 mm (7 x 13.5 x 17 inches) (height x width x depth) | |
Weight | 17 kg (38 lbs) |